Device Interfacing & Integration Policy

Product: Caresoft eICU  •  Version: 1.1  •  Effective: 01/04/2026

This Policy governs how bedside medical devices are connected to the Caresoft eICU Platform, who is responsible for what, and what must be validated before any device data is used clinically. It forms part of the Master Services Agreement and must be read with the Intended Use & Clinical Safety Statement.

THE INTERFACE IS READ-ONLY IN BOTH DIRECTION AND INTENT.

The Platform receives data from devices. It does not send commands, change settings, silence alarms, alter alarm limits, or influence device behaviour in any way. No feature that writes to a bedside device will be introduced without a separate, formal regulatory and clinical assessment.

Connecting a device to the Platform must not alter that device's configuration, alarm behaviour, or its own regulatory status.

Contents
  1. Scope
  2. Validated device scope
  3. Unvalidated devices
  4. Interface architecture
  5. Network and infrastructure requirements
  6. Responsibility matrix
  7. Validation protocol
  8. Site acceptance testing
  9. Time synchronisation
  10. Bed, device and patient attribution
  11. Integrity monitoring
  12. Data gaps and quarantine
  13. Change control and re-validation
  14. Biomedical and vendor coordination
  15. Security of the device network
  16. Decommissioning
  17. Records to be maintained
  18. Contact

1. Scope

This Policy applies to every bedside device connected to the Platform at every deploying site — patient monitors, ventilators, anaesthesia machines, infusion devices and any other source of physiological data — and to the edge agents, gateways, cabling and network segments used to carry that data.

2. Validated device scope

Only devices that appear on the site-specific validated device list may be interfaced and relied upon clinically. A device qualifies for that list only when the specific make, model, software/firmware version and interface mode has passed bench validation and site acceptance testing under Sections 7 and 8.

Current product-level supported scope:

Vendor / familyInterfaceDriverStatus
Philips IntelliVueHL7 ORU^R01 over MLLPHL7/MLLP + vendor profile[Validated / pending — state per model & firmware]
GE CARESCAPEHL7 ORU^R01 over MLLPHL7/MLLP + vendor profile[…]
Mindray BeneVisionHL7 ORU^R01 over MLLPHL7/MLLP + vendor profile[…]
Nihon KohdenHL7 ORU^R01 over MLLPHL7/MLLP + vendor profile[…]
Dräger ventilators / anaesthesiaMedibusMedibus driver[…]
BPL, Schiller, Skanray, legacy serial devicesASCII over serialASCII serial driver[…]

Vendor support for a protocol is not validation. A vendor family appearing above means a driver exists; it does not mean any particular unit in your ICU is validated. Validation is per model, per firmware version, per site.

Explicitly out of Phase 1 scope: Philips DEI over RS-232, waveform capture, and any device requiring a proprietary licence or gateway not procured by the hospital.

3. Unvalidated devices

Data from a device not on the site's validated list must not be displayed to clinical users or relied upon for any clinical purpose.

Where an unvalidated device is connected for testing, it must be confined to a non-clinical environment or clearly flagged as unvalidated in the Platform, and clinical users must be informed in writing that it is not to be used.

Connecting an unvalidated device to the production Platform without written agreement is a breach of this Policy and of the Master Services Agreement, and voids Caresoft's representations in respect of the affected data.

4. Interface architecture

5. Network and infrastructure requirements

The hospital must provide and maintain:

6. Responsibility matrix

R = responsible   S = supports   A = accountable/approves

ActivityHospital
clinical
Hospital
biomed / IT
Device
vendor
Caresoft
Device ownership, calibration, preventive maintenance, electrical safetyR/AS
Enabling and licensing the device data output port/protocolRRS
Supplying protocol specification and conformance statementSRS
Providing device, cabling, adapters, network and edge hardwareR/AS
Driver development and parameter mappingSSR
Bench validation of driver against deviceSSR
Site acceptance test executionSRR
Site acceptance test approvalAAS
Maintaining the validated device list for the siteSRR
Bed–device mapping and keeping it currentSR/AS
Accurate and timely ADT informationR/ASS
Notifying Caresoft before firmware upgrade or device replacementR/AS
Setting clinical notification thresholdsR/AS
Clinical interpretation and all clinical decisionsR/A
Platform availability, ingest, storage, auditR/A
Monitoring sync health and acting on integrity alertsSRR
Incident reporting under the Intended Use Statement §17RRR

Where the hospital has no in-house biomedical engineering function, it must name the party performing that role — an external biomedical service provider or the device vendor — in writing before go-live. These responsibilities cannot be left unassigned, and Caresoft does not assume them by default.

7. Validation protocol

Each device make/model/firmware/interface combination is validated against a documented protocol before it may be added to any site's validated list. The protocol covers at minimum:

  1. Procurement and documentation — protocol specification, conformance statement, licence for the data output where required.
  2. Connection — physical and logical connection established without altering device configuration.
  3. Parameter identification — every parameter mapped to the correct internal code, with units confirmed. Unit mismatches and vendor-specific codes explicitly checked.
  4. Value accuracy — displayed device value compared against Platform-received value across the clinical range, including boundary and out-of-range values.
  5. Special values — handling of null, invalid, artefact-flagged, out-of-range, "searching", and probe-off states verified. These must not be presented as valid numbers.
  6. Timestamp fidelity — device time versus received time, drift measured.
  7. Sampling behaviour — interval, jitter and burst behaviour characterised.
  8. Disconnection — behaviour on cable pull, device power-off, network loss and device standby; confirm the Platform shows a gap rather than a stale value.
  9. Reconnection and replay — buffered data uploads correctly with original timestamps, no duplication, no reordering error.
  10. Load — sustained operation at expected bed count for a defined period without degradation.
  11. Alarm independence — confirm device alarms are unaffected by connection and continue to function normally.
  12. Negative testing — malformed messages, wrong signature, wrong device identity, clock skew beyond tolerance — all rejected and logged, none silently accepted.

Results are recorded in a signed validation report retained by both parties.

8. Site acceptance testing

9. Time synchronisation

Incorrect time is a patient safety issue. It corrupts trends, breaks attribution across transfers, and can make the sequence of clinical events unreconstructable after an incident.

10. Bed, device and patient attribution

11. Integrity monitoring

The Platform records and surfaces its own integrity state. The hospital must assign responsibility for reviewing it:

SignalMeaningReview
Sync health per deviceWhether data is currently flowingContinuous — command centre
Missed heartbeatsEdge agent or gateway downImmediate — hospital IT
Clock offsetTime drift beyond toleranceDaily — hospital IT
Data gapsPeriod with no data for a bedPer shift — clinical
Quarantined samplesAttribution failureDaily — client admin
Rejected batchesSignature or format failureDaily — Caresoft + hospital IT
Error logDriver and ingest errorsDaily — Caresoft

A device showing no data is not a patient with normal physiology. Clinical users must check sync health before interpreting any absence of data, and must never treat a flat or empty trend as reassurance.

12. Data gaps and quarantine

13. Change control and re-validation

The hospital must notify Caresoft in writing at least [14] days before any of the following. Proceeding without notice may cause silent data loss or corruption that neither party detects until an incident.

ChangeRe-validation required
Device firmware or software upgradeYes — full parameter and value verification
Device replacement, same model and firmwareAbbreviated SAT — connection, identity, parameter check
New device model or new vendorYes — full bench validation and SAT
Change of interface mode or protocol settingsYes
Device moved between bedsNo re-validation; bed–device mapping must be updated
Network re-architecture, VLAN or firewall changeConnectivity and load re-test
Edge agent hardware or OS changeConnectivity, buffering and clock re-test
Platform release affecting drivers, mapping, scoring or attributionPer Intended Use §18
New ICU, new ward or bed count expansionSAT for the new area; licence check

Where a change is made without notice and data integrity is affected, remediation is chargeable and Caresoft's representations in respect of the affected data do not apply.

14. Biomedical and vendor coordination

15. Security of the device network

16. Decommissioning

17. Records to be maintained

Both parties retain, for the life of the deployment plus the retention period in the Master Services Agreement:

18. Contact

Device integration: [email protected]
Patient safety incidents: [email protected] and 7400390415 — immediate
Security: [email protected]
Caresoft Systems Private Limited, 311, Mahesh Industrial Estate , Silver Park, Mira Road East , Thane -401107,
CIN U72900MH2022PTC387875.

Home Book a Demo